Density estimation based on the Hounsfield unit value of cone beam computed tomography imaging of the jawbone system

The aim of this study is to investigate the estimation of density from the Hounsfield unit of cone beam computed tomography data in dental imaging, especially for dental implant application. A jaw phantom with various known densities of anatomical parts (e.g. soft tissue...

詳細記述

書誌詳細
出版年:Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
第一著者: 2-s2.0-85058694413
フォーマット: 論文
言語:English
出版事項: SAGE Publications Ltd 2018
オンライン・アクセス:https://www.scopus.com/inward/record.uri?eid=2-s2.0-85058694413&doi=10.1177%2f0954411918806333&partnerID=40&md5=f98a8abf14363926d2d268eff7369e7c
その他の書誌記述
要約:The aim of this study is to investigate the estimation of density from the Hounsfield unit of cone beam computed tomography data in dental imaging, especially for dental implant application. A jaw phantom with various known densities of anatomical parts (e.g. soft tissue, cortical bone, trabecular bone, tooth enamel, tooth dentin, sinus cavity, spinal cord and spinal disc) has been used to test the accuracy of the Hounsfield unit of cone beam computed tomography in estimating the mechanical density (true density). The Hounsfield unit of cone beam computed tomography data was evaluated via the MIMICS software using both two-dimensional and three-dimensional methods, and the results showed correlation with the true density of the object. In addition, the results revealed that the Hounsfield unit of cone beam computed tomography and bone density had a logarithmic relation, rather than a linear one. To this end, the correlation coefficient of logarithmic correlation (R 2 = 0.95) is higher than the linear one (R 2 = 0.77). © IMechE 2018.
ISSN:9544119
DOI:10.1177/0954411918806333