High flex femoral component with slanted pegs improves fixation strength of total knee arthroplasty

High incidences of early aseptic loosening cases have led to numerous analyses on fixation strength of femoral component (FC). This study consists of two objectives: to analyze the correlation between articulating contact area shape conformities and fixation strength, and (2) to analyze the relation...

وصف كامل

التفاصيل البيبلوغرافية
الحاوية / القاعدة:Materials Today: Proceedings
المؤلف الرئيسي: Arsat Z.A.; Uda M.N.A.; Abdullah F.; Hashim M.K.R.; Mutallib M.F.A.; Kadarman A.H.; Ahmed Shokri A.; Aziz M.E.; Shuib S.
التنسيق: مقال
اللغة:English
منشور في: Elsevier Ltd 2023
الوصول للمادة أونلاين:https://www.scopus.com/inward/record.uri?eid=2-s2.0-85147026566&doi=10.1016%2fj.matpr.2023.01.098&partnerID=40&md5=e4ea7609349a0d7322c27b479e799c5b
الوصف
الملخص:High incidences of early aseptic loosening cases have led to numerous analyses on fixation strength of femoral component (FC). This study consists of two objectives: to analyze the correlation between articulating contact area shape conformities and fixation strength, and (2) to analyze the relationship between slanting box patterns of the FC and fixation strength. Two design analyses were constructed: first, using three different types of contact congruence with radius ratios of 0.50, 0.89 and flat. Second, box patterns of the FC were anteriorly slanted at 2, 4 and 6°. The fixation strength of the FC is significantly improved (38.59%) by slanting the box pattern anteriorly as compared to the non-slanted FC design. © 2023 Elsevier Ltd. All rights reserved.
تدمد:22147853
DOI:10.1016/j.matpr.2023.01.098