Monte carlo dose calculation in dental amalgam phantom

It has become a great challenge in the modern radiation treatment to ensure the accuracy of treatment delivery in electron beam therapy. Tissue inhomogeneity has become one of the factors for accurate dose calculation, and this requires complex algorithm calculation like Monte Carlo (MC). On the oth...

Full description

Bibliographic Details
Published in:Journal of Medical Physics
Main Author: Aziz M.Z.A.; Yusoff A.L.; Osman N.D.; Abdullah R.; Rabaie N.A.; Salikin M.S.
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2015
Online Access:https://www.scopus.com/inward/record.uri?eid=2-s2.0-84941660916&doi=10.4103%2f0971-6203.165080&partnerID=40&md5=0f41cd8bd151e053fe02982e88022903
Description
Summary:It has become a great challenge in the modern radiation treatment to ensure the accuracy of treatment delivery in electron beam therapy. Tissue inhomogeneity has become one of the factors for accurate dose calculation, and this requires complex algorithm calculation like Monte Carlo (MC). On the other hand, computed tomography (CT) images used in treatment planning system need to be trustful as they are the input in radiotherapy treatment. However, with the presence of metal amalgam in treatment volume, the CT images input showed prominent streak artefact, thus, contributed sources of error. Hence, metal amalgam phantom often creates streak artifacts, which cause an error in the dose calculation. Thus, a streak artifact reduction technique was applied to correct the images, and as a result, better images were observed in terms of structure delineation and density assigning. Furthermore, the amalgam density data were corrected to provide amalgam voxel with accurate density value. As for the errors of dose uncertainties due to metal amalgam, they were reduced from 46% to as low as 2% at d80 (depth of the 80% dose beyond Zmax) using the presented strategies. Considering the number of vital and radiosensitive organs in the head and the neck regions, this correction strategy is suggested in reducing calculation uncertainties through MC calculation.
ISSN:9716203
DOI:10.4103/0971-6203.165080