Electrochemical sensor for the lead determination: Development and characterisation

Lead electrochemistry sensor had been developed using screen printed electrode (SPE) technique with poly(vinyl) chloride (PVC) as printed matrix. The study has been carried out by using different type of electrodes. However, lead electrochemistry sensor had optimum response when deposited with mercu...

Full description

Bibliographic Details
Published in:2005 Asian Conference on Sensors and the International Conference on New Techniques in Pharmaceutical and Biomedical Research - Proceedings
Main Author: Loh H.C.; Ahmad M.; Taib M.N.
Format: Conference paper
Language:English
Published: 2005
Online Access:https://www.scopus.com/inward/record.uri?eid=2-s2.0-33846298895&doi=10.1109%2fASENSE.2005.1564520&partnerID=40&md5=ccec350cc1f03571c81f435d49be3dd7
Description
Summary:Lead electrochemistry sensor had been developed using screen printed electrode (SPE) technique with poly(vinyl) chloride (PVC) as printed matrix. The study has been carried out by using different type of electrodes. However, lead electrochemistry sensor had optimum response when deposited with mercury on the top of working electrode and in the buffer solution of pH 2. The optimum parameters for mercury deposition include deposited time = 15 s, deposited potential = 1000 mV, scanning range = 20 mV/s and amplitude pulse = 25. This sensor has linearity range of 1 - 50 ppm of Pb(II) R2 of 0.9997. The LOD of this sensor is lower than 1 ppm of Pb(II). ©2005 IEEE.
ISSN: